Real device photo with interactive controls — the same UFD 900 used in my Udemy courses UT Level 1 and UT Level 2.
In this trial version you calibrate (adjust) the probe on a step block, then find flaws in a steel plate and identify an unknown material by ultrasound. Use the Tour function to be guided step by step.
Position: 150 mm
Single-element straight-beam probe.
Drag the probe along the plate's surface. Click it to lift it off (line goes flat) or place it back down (coupled).
Press Power first. The 5 tiles on the right of the display are real
menus, exactly like the original device: Range, Probe,
Gate, Set and Mem jump straight into their own menu
(BASE/PROBE/GATE/SET/MEM), DAC/AVG cycles DAC → AVG → off. The
Menu key (between dB+ and ↑) opens the grid of function groups — move
the cursor with the D-pad, Enter loads a group,
Cancel closes it without changing anything.
Inside a menu, tapping a tile selects it (pink outline) or fires the
action, and ↑/↓ step through a menu's own functions, scrolling if it
has more than 5. Change a value with the rotary encoder (click
left/right or scroll wheel) or the ◀/▶ arrows. dB+/dB− change
gain directly, even outside the GAIN menu. The round key with the
zigzag icon is AutoGain. The second square icon is Zoom.
Freeze shows "Freeze" on screen. On the test specimen to the
right, drag the probe along the plate's surface or click it to lift it
off / place it back down — without coupling the trace stays flat. A
hidden flaw sits somewhere along the plate: scan slowly and watch for
an extra echo ahead of the first backwall signal — like a real
near-surface reflector, it rings its own repeat-echo train too
(fading faster than the backwall's), so don't mistake those for
deeper flaws.
Picking a material here dials MATVEL to its reference velocity, same as typing it in on the real device. Every test specimen is steel (5920 m/s) except the DAC Block (aluminium, 6360 m/s — no mystery, it’s labeled on the block) and the Unknown Material block, whose real velocity is deliberately hidden — its known 30 mm thickness only reads correctly once MATVEL is right. Wrong MATVEL = wrong depth reading everywhere, just like on the real device.
Angle beam probe — refraction angle
Pick the angle stamped on your probe (ANGLE in the PROBE menu). The V2-Normal block carries a scale for checking that angle against the Ø5 hole; it runs from 35° to 65°, numbered at 40/50/60/65 like the real block — so a 70° or 75° probe has no mark to index against there.
Crystal size (rectangular element)
The element's real width×height, stamped on the probe housing — a bigger crystal narrows the beam's divergence and raises its near-field length.
Nominal frequency
Also stamped on the probe housing (X_FREQ in the PROBE menu — same setting, just picked here too since it's part of the same probe ID).
SET files
Please confirm
Not available
Enter name
Report
Test Result
Parameters
Date & Time
Read-only in this UT-Trainer — the instrument's clock always follows this device's own system clock.
DEV — couplant line geometry
Live-adjusts the yellow couplant line. Nothing here is visible to a normal viewer — sliders only, no numbers move anything until you drag one.
Applies to
Sliders below load from — and on OK, write into — every checked probe. Each one starts with its own default; check more than one to make them share a value.
Live-trace jitter ("Zappeln")
Sound attenuation — typical values
Typical attenuation of the longitudinal wave in dB/m at three probe frequencies — the values the UT-Trainer uses for its specimens (source: K. Ono, “A Comprehensive Report on Ultrasonic Attenuation of Engineering Materials”, Applied Sciences 10 (2020) 2230; aluminium matched to a real 30 mm block). Attenuation rises with frequency and grain size; for transverse waves (angle probe) the trainer uses twice the value. Real materials differ widely with structure and heat treatment, so for accurate work the attenuation is measured on the test piece itself. For the AVG evaluation enter the value under “Sound attenuation” (CORRECT box or EDIT AVG tile).
Sound velocities — common materials
Typical longitudinal- and transverse-wave velocities. Where the usual reference tables (Olympus/Evident, NDT Resource Center, Onda, Waygate Krautkrämer, Dakota) differ by more than about 2 %, the range is given — mostly because of alloy, heat treatment or, for cast iron, the graphite form. Liquids carry no transverse waves. For accurate measurements set VELOCITY on a reference block of the same material with known thickness, as with the Unknown Material specimen.